Abstract

In order to design the prosthesis which can mimic the surroundings for bone cell proliferation and cell attachment and avoid stress shielding effect by matching the Young’s modulus with different parts of human bone, we come up with the porous structure with scaffold design. This paper is presenting the scaffold of design, the mechanical property of the biomimetic scaffold and the relative error rate between the Finite element analysis (FEA) and compressive test. However, most of the models of the porous structure are designed in the STL format especially the high topological model which will make the FEM preprocessing procedure difficult. According to the reason above, we used the concept of the CT-based voxel mesh method for high complex STL model, which skips the time-consuming mesh and repairing reconstruction process. Lastly, we successfully designed the porous prosthesis according to the data from FEA and compressive test.