Abstract

The report describes a method for the automated polarographic determination of cyanide as tetracyanonickelate (II) anion complex in a gas-diffusion flow system. The volatile cyanide, existing in whole blood, plasma and urine samples, was measured after gas-diffusion using 8 x 10-5 mol l-1 hexaaminenickel solution as acceptor. The linear range of calibration, for measurements at the hanging mercury-drop electrode (HMDE), was from 0.1 to 2.0 micrograms cyanide with r = 0.998. The RSD was, respectively, 3.4 and 1.2% (n = 5) for 0.4 microgram cyanide measured with and without the flow-system configuration. Detection limits of 7.4 microgram l-1 were calculated using the flow system and the method was compared with the classical method using Cavet flasks. Parameters that affect the cyanide determination in the proposed method, such as acceptor solution, pH, flow rate and temperature, were investigated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.