Abstract

A method for the polarographic determination of cyanide as contaminant in pralidoxime mesylate (PM) formulations was developed. The volatile cyanide formed in the formulations was stabilized as tetracyanonickelate (TCN) anion complex after reaction with ammoniacal Ni(II) solution. The stable TCN anion complex ( K stb=10 31) was determined by anodic stripping voltammetry at the hanging mercury drop electrode (HMDE). The polarographic signal was proportional to the cyanide concentration and the high concentration of PM did not interfere. The linear range of calibration was from 1.2 to 16 μg cyanide with r=0.998. The RSD was 1.3% ( n=5) for 2.4 μg cyanide and a detection limit of 0.8 μg cyanide was calculated. The proposed method is adequate as a quality control of PM formulations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.