Abstract

A coupled ion chromatography (IC) system was used for the determination of chloride, sulphate and phosphate in high-purity nitric acid. Such a high ionic strength matrix causes a selectivity problem in single IC systems. The first part of the system is used for a pre-separation of the analytes from the nitrate matrix. A specially designed high-capacity anion exchanger with low retention for the analytes and high retention for nitrate was developed. The eluent stream containing the analytes was transferred to the second part of the system via a heart-cut valve and a pre-concentration column. The second system utilizes a high performance anion exchanger and is used to quantify the analytes. Recoveries of the analytes are 80–100% for phosphate, and around 100% for sulphate and chloride. Detection limits for chloride, sulphate and phosphate in concentrated nitric acid (69% w/w) are 0.1, 1 and 5 mg/l, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.