Abstract

Abstract Theophylline, a bronchodilator used in the treatment of respiratory diseases, displays a narrow therapeutic range requiring therapeutic drug monitoring to maximize its effectiveness and safety. Current measurements of theophylline concentrations are, however, achieved using chromatography or immunoassays, which, in spite of their high accuracy and sensitivity, rely on expensive laboratory-based instruments operated by trained personnel or provide only semi-quantitative results, respectively. In response, we developed an electrochemical DNA aptamer-based (E-AB) sensor for theophylline on screen-printed electrodes. Our sensors enable rapid (<30 s) and selective measurements against theophylline structural analogs in clinically relevant range in finger-pricked-sized volumes of undiluted blood samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.