Abstract
AbstractA technique is proposed which allows introduction of very large volumes of liquid (10 ml were tested) into capillary columns equipped with short (1–2 m long) retention gaps. It is based on concurrent solvent evaporation, i.e. evaporation of the solvent during introduction of the sample. The technique presupposes high carrier gas flow rates (at least during sample introduction) and column temperatures near the solvent boiling point. The major limitation of the method is the occurrence of peak broadening for solutes eluted up to 30°, in some cases up to 100°, above the injection temperature. This is due to the absence of solvent trapping and a reduced efficiency of phase soaking. Therefore, use of volatile solvents is often advantageous. Application of the concurrent solvent evaporation technique allows introduction of liquids which do not wet the retention gap surface. However, the method is still not very attractive for analysis of aqueous or water‐containing solutions (reversed phase HPLC).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.