Abstract

This study compared the biomechanical behaviour of repairs in the human flexor digitorum profundus tendon in zones I, II and III with repairs of different segments of the porcine flexor tendon of the second digit and the extensor digiti quarti proprius tendon, in order to assess the validity of porcine tendons as models for human flexor tendon repairs. These porcine tendons were selected after comparing their size with the human flexor digitorum profundus tendon. The tendon repairs were done in three segments of each porcine tendon and repairs in the human tendons were done in zones I,II and III. Ten tendons in each group yielded a total of 90 specimens. A modified Kessler repair was done with 3-0 coated braided polyester suture and subjected to uniaxial tensile testing. In human flexor tendons, the ultimate force was higher in zones I and II than in zone III. The porcine flexor digitorum profundus tendon from the second digit and the proximal segment of the extensor digiti quarti proprius tendon behaved similarly to the human flexor tendon in zone III and can be considered as surrogates for the human flexor tendon.