Abstract

BackgroundA Y-shaped rotatable connector (YRC) for double-lumen tubes (DLT) is invented and compared with the traditional connector (Y-shaped connector, YC).MethodsSixty patients with ASA grade I-III, aged ≥ 18 years, who needed to insert a DLT for thoracic surgery were recruited and assigned into the YRC group (n = 30) and the YC group (n = 30) randomly. The primary endpoints included the inhaled air concentration (Fi) and the exhaled air concentration (Et) of sevoflurane before and after the switch between two-lung ventilation and one-lung ventilation at different times, positioning time, and switching time. The secondary endpoints were the internal gas volume of the two connectors, airway pressure, and the sputum suction time.ResultsThe Et and Fi of the YRC group and the YC group were significantly different (all p < 0.05) at 5s, 10s, and 30s after the patient switched from two-lung ventilation to one-lung ventilation. The positioning time of the YRC group was less than YC group (89.75 ± 14.28 s vs 107.80 ± 14.96 s, p < 0.05), as well as the switching time (3.60 ± 1.20 s vs 9.05 ± 2.53 s, p < 0.05) and the internal gas volume (17.20 ml vs 24.12 ml). There was no difference in airway pressure and the sputum suction time in two groups.ConclusionCompared with YC, YRC was beneficial for maintaining depth of anesthesia, improves efficiency for the switch between one-lung and two-lung ventilation, and shortens the tube positioning time.