Abstract
The results of studying the band characteristics and the efficiency of recording the secondary-electron multipliers VEU-6 and VEU-GOI (manufactured by the Vavilov State Optical Institute (GOI), St. Petersburg) within the ultrasoft X-ray range are given. The measurements were carried out at the metrological station Kosmos with the use of synchrotron radiation from the storage ring VEPP-4M (Budker Institute of Nuclear Physics, Siberian Branch, Russian Academy of Sciences, Novosibirsk) in monochromatic radiation at a photon energy of 268 eV. A strong irregularity in the band characteristics of both VEU-6 and VEU-GOI has been found. Their quantum efficiencies are estimated. The measurements are a continuation of the works on calibration of the equipment of the Space Solar Patrol (developed by GOI) using synchrotron radiation from the VEPP-4M storage ring (project MNTTs no. 2500).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Surface Investigation. X-ray, Synchrotron and Neutron Techniques
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.