We have constructed a simple positron beam for measuring both the efficiencies of positron moderator foils and the positron yields of encapsulated positron beam sources. The slow-positron emission rate from the moderator is determined from the positron annihilation radiation. This technique is also applied to measure source yields. In addition to the coincidence technique, the source yield is determined by measuring electric currents generated by the annihilating positrons. The performance of the system is demonstrated with different tungsten moderator foils before and after heat treatments, and several Na 22 sources with primary activities in the range 1–50 mCi.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.