Abstract
We have constructed a simple positron beam for measuring both the efficiencies of positron moderator foils and the positron yields of encapsulated positron beam sources. The slow-positron emission rate from the moderator is determined from the positron annihilation radiation. This technique is also applied to measure source yields. In addition to the coincidence technique, the source yield is determined by measuring electric currents generated by the annihilating positrons. The performance of the system is demonstrated with different tungsten moderator foils before and after heat treatments, and several Na 22 sources with primary activities in the range 1–50 mCi.
Paper version not known (Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have