Abstract

Various diagnostic techniques have been applied at the neutron-induced positron source Munich NEPOMUC in order to determine the positron beam parameters such as intensity, beam shape and energy distribution. The positron beam intensity is determined by the detection of the annihilation radiation of positrons, which annihilate in a movable target. The use of a micro-channel plate (MCP) detector with a CCD-camera allows a direct measurement of the positron beam shape and the lateral resolved intensity distribution. At NEPOMUC a movable MCP-assembly inside the evacuated beam line enables a quick examination of the beam shape during operation. A retarding grid was mounted inside the homogeneous magnetic guiding field in order to determine the distribution of the longitudinal positron momentum, and hence estimate the energy distribution of the positrons.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.