Abstract

One-step immunochromatographic assay (ICA) has been developed using colloidal gold-labeled monoclonal antibody probe for the rapid detection of lead ions in water samples. The ICA was based on the theory of competitive reactivity, and the results can be easily judged based on the presence or absence of a red colored test line with visual detection. Under optimal conditions, this method shows high detecting sensitivity with a LOD (limit of detection) of 50 ng/ml. Stability test indicates that the immunochromatographic strips are stable for 8 weeks at room temperature. During practical application, nanometer TiO 2 is used to enrich the lead ions in water samples. The ICA is successfully applied in the measurement of lead ion concentrations in local water samples, and the results are highly consistent with that of ICP-MS. Detecting lead ions with ICA can be done within 4 min and is very useful for the rapid onsite testing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.