Abstract

In this work, overshoot and undershoot effects associated with step‐and‐shoot IMRT (SSIMRT) delivery on a Varian Clinac 21iX are investigated, and their impact on patient‐specific QA point dose measurements and treatment plan delivery are evaluated. Pinnacle3 SSIMRT plans consisting of 5, 10, and 15 identical 5×5 cm2 MLC defined segments and MU/segment values of 5 MU, 10 MU, and 20 MU were utilized and delivered at 600/300 MU/min. Independent of the number of segments the overshoot and undershoot at 600 MU/min were approximately ±10%,±5%, and ±2.5% for 5 MU/segment, 10 MU/segment, and 20 MU/segment, respectively. At 300 MU/min, each of these values is approximately halved. Interfractional variation of these effects (10 fractions), as well as dosimetric variations for intermediate segments, are reduced at the lower dose rate. QA point‐dose measurements for a sample (n=29) of head and neck SSIMRT beams were on average 2.9% (600 MU/min) and 1.7% (300 MU/min) higher than Pinnacle3 planned doses. In comparison for prostate beams (n=46), measured point doses were 0.8% (600 MU/min) and 0.4% (300 MU/min) higher. The reduction in planned‐measured point‐dose discrepancies at 300 MU/min can be attributed in part to the inclusion of the first segment (overshoot) in the admixture of segments that deliver measured dose. Pinnacle3 plans for 10/9 head and neck/prostate treatments were adjusted by ±0.5 MU to include the effects of overshoot and undershoot at 600 MU/min. Comparing original and adjusted plans for each site indicated that the original plan was preferred in 70% and 89% of head and neck and prostate cases, respectively. The disparity between planned and delivered treatment that this suggests can potentially be mitigated by treating SSIMRT at a dose rate below 600 MU/min.PACS number(s): 87.55.Qr, 87.56.bd, 87.56.N‐

Highlights

  • 115 Baines et al.: Clinical implications of MLC overshoot tracks the cumulative MU for a given beam and initiates beam termination when the planned MU is delivered.(5) the overshoot, coupled with less significant MU discrepancies for intermediate segments,(4) causes the last beamlet to be interrupted before the planned MU is delivered, the so-called undershoot effect

  • Interfractional variations of overshoot and undershoot at both dose rates are shown in Fig. 2 for a 10-segment 5 MU/ segment step-and-shoot IMRT (SSIMRT) beam

  • A delayed beam-on coupled with a delayed beam-off give rise to Δti values that are generally less than those of the first and last segments.(2) For the last segment, beam delivery is terminated when the linac monitor identifies that the cumulative MU for the SSIMRT beam has been achieved

Read more

Summary

Introduction

115 Baines et al.: Clinical implications of MLC overshoot tracks the cumulative MU for a given beam and initiates beam termination when the planned MU is delivered.(5) the overshoot, coupled with less significant MU discrepancies for intermediate segments,(4) causes the last beamlet to be interrupted before the planned MU is delivered, the so-called undershoot effect. The significance of these effects increases with decreasing MU/segment and decreases with decreasing dose rate.(3) For a given dose rate, the magnitude of the overdose and underdose are approximately the same. The dose rate dependence of these effects was evaluated using dose rates of 600 MU/min (default value at our facility) and 300 MU/min

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call