Abstract

Single plan techniques for multiple brain targets (MBT) stereotactic radiosurgery (SRS) are now routine. Patient specific quality assurance (QA) for MBT poses challenges due to the limited capabilities of existing QA tools which necessitates several plan redeliveries. This study sought to develop an SRS QA phantom that enables flexible MBT patient specific QA in a single delivery, along with complex SRS commissioning. PLA marble and PLA StoneFil materials were selected based on the literature and previous research conducted in our department. The HU numbers were investigated to determine the appropriate percentage infill for skull and soft-tissue equivalence. A Prusa MK3S printer in conjunction with the above-mentioned filaments were used to print the SRS QA phantom. Quality control (QC) was performed on the printed skull, film inserts and plugs for point dose measurements. EBT3 film and point dose measurements were performed using a CC04 ionisation chamber. QC demonstrated that the SRS QA phantom transverse, coronal and sagittal film planes were orthogonal within 0.5°. HU numbers for the skull, film inserts and plugs were 858 ± 20 and 35 ± 12 respectively. Point and EBT3 film dose measurements were within 2.5% and 3%/2mm 95% gamma pass rate, respectively except one Gross Tumour Volume (GTV) that had a slightly lower gamma pass rate. Dose distributions to five GTVs were measured with EBT3 film in a single plan delivery on CyberKnife. In conclusion, an SRS QA phantom was designed, and 3D printed and its use for performing complex MBT patient specific QA in a single delivery was demonstrated.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call