Abstract
We report measurements of chemical concentrations in clinical blood serum and urine samples using liquid-core optical fiber (LCOF) Raman spectroscopy to increase the collected signal strength. Both Raman and absorption spectra were acquired in the near-infrared region using the LCOF geometry. Spectra of 71 blood serum and 61 urine samples were regressed via partial least squares against reference analyzer values. Significant correlation was found between predicted and reference concentrations for 13 chemicals. Using absorption data to normalize the LCOF enhancement made the results more accurate. The experimental geometry is well suited for high-volume and automated chemical analysis of clear biofluids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.