Abstract

LiF:Mg,Cu,P phosphors (commercially designated as MCP-N) are very sensitive thermoluminescent detectors and are considered by many laboratories to hold great promise for low level radiation monitoring. It is their high efficiency relative to TLDs based on LiF:Mg,Ti which make MCP-N particularly attractive for low dose measurements. The currently investigated detector material made of LiF:Mg,Cu,P bears the code name MCP-7s and has a very thin active layer and consequently an improved response to low energy beta radiation. Detailed studies have been carried out to characterise MCP-7s detectors in terms of glow curve, beta and gamma response, reusability, batch homogeneity and fading. Based on the positive outcome of these investigations, the dosimetry service at PSI plans to replace the TL material LiF:Mg,Ti presently used for extremity dosimetry with MCP-7s

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.