Abstract
C-D and A-T phantoms were imaged using a range of clinical breast ultrasound systems and imaging modes. A novel sensitive imaging performance metric, the Detectability Score (DS), was proposed which encompasses the Lesion Contrast to Noise Ratio (LCNR) weighted by the lesion depth and diameter. A geometry-based theoretical model comparing LCNR measured using spherical and cylindrical phantom anechoic/lesion targets was developed to investigate the influence of slice thickness on focal lesion detectability. LCNR and DS metrics derived from phantom image measurments were capable of differentiating the imaging performance of a range of ultrasound systems and advanced imaging modes, with the -2 dB contrast lesion targets offered as the most challenging to resolve. The geometry-based theoretical model, validated against phantom measurements, demonstrated the significant influence of slice thickness on focal lesion detectability, highlighting the need for increased availability of low contrast resolution spherical target phantoms for clinically realistic performance evaluation. The performance metrics coupled with the -1 dB contrast targets provide scope for evaluating future technological improvements in ultrasound systems. Given the high dependence of breast cancer care on high quality ultrasound imaging techniques, there is a need for evaluating imaging performance using clinically relevant test objects.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have