Abstract

Chitosan (CS) is widely used as a scaffold material in tissue engineering. The objective of this study was to test whether porous chitosan membrane (PCSM) coating for Nafion used in implantable sensor reduced fibrous capsule (FC) density and promoted superior vascularization compared with PCSM coating for polytetrafluoroethylene (PTFE). PCSM was fabricated with solvent casting/particulate leaching method using silica gel as porogen and characterized in vitro. Then, PCSM-Nafion and PCSM-PTFE composites were assembled with hydrated PCSM and implanted subcutaneously in rats. The histological analysis was performed in comparison with Nafion and PTFE. Implants were explanted 35, 65, and 100 days after the implantation. Histological assessments indicated that both composites achieved presumed effects of porous coatings on decreasing collagen deposition and promoting angiogenesis. PCSM-PTFE exerted higher collagen deposition by area ratio, both within and outside, compared with that of PCSM-Nafion. Angiogenesis within and outside the PCSM-Nafion both increased over time, but that of the PCSM-PTFE within decreased.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.