Abstract
An automated high-performance liquid chromatographic column-switching system is presented for the analysis of antiepileptic drugs in serum. Initial results show that a reversed-phase extraction column works best overall when fitted with screens versus frits, and when packed with porous 30-μm particles as opposed to a pellicular packing of similar size or with smaller porous particles. The continuous analysis of primidone for over 2000 serum samples is achieved at a rate of twelve samples per hour with a single analytical column. An analogous boxcar high-performance liquid chromatographic system is also assembled and used to analyze two of four injected antiepileptic drugs at a rate of 40 samples per hour. For 1000 of these analyses, the coefficient of variation is 1% without an internal standard.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.