Abstract

We have developed a semi-automated method for measuring five kinds of antiepileptic drugs in serum by successfully adapting commercial competitive-binding enzyme immunoassay kits (MARKIT; Dainippon) for use with a continuous-flow analyzer (Technicon AutoAnalyzer II equipped with a dialyzer). The free enzyme-labeled drug is automatically separated by a microfilter from the competitive immunoreaction mixture between labeled and unlabeled drug for anti-drug immunoglobulin coupled to bacterial cell walls. The concentrations of the antiepileptic drugs in serum samples can be determined by automated measurement of enzyme activity of the enzyme-labeled drugs. Results of the semi-automated method correlated well with those obtained by manual enzyme immunoassay, gas-liquid chromatography, and "high-pressure" liquid chromatography. The correlation coefficients were all greater than 0.95, showing the practicality of this method for therapeutic monitoring of antiepileptic drugs.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call