Abstract

We compared creatinine concentrations in serum and urine and creatinine clearances determined by two Jaffé (Beckman's "Astra," Boehringer Mannheim Diagnostics) and two enzymatic (Kodak, Boehringer Mannheim Diagnostics) methods. Serum creatinine and creatinine clearances determined by each method were also compared with the glomerular filtration rate as measured with use of sodium [125I]iothalamate in patients with a wide range of renal function. Results between methods correlated excellently, but we saw clear method-dependent biases of up to 2.9 mg/L for serum. The highest serum creatinine values and the lowest creatinine clearances were obtained with Boehringer Mannheim Diagnostics' Jaffé method. The reciprocal of the serum creatinine and the creatinine clearance also correlated well with the glomerular filtration rate, but all methods over-estimated the glomerular filtration rates to varying degrees. Appropriate standardization of methods appears to be as important as method principle for establishing an accurate relationship between creatinine determinations and glomerular filtration rate.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.