Abstract
The current research was done to assess microbial seepage of two types of internal implant-abutment connections. Twenty dental implants are categorized into two groups. Group A fixtures with an internal hexagonal geometry and group B fixtures with a tri-lobe internal connection. All implant-abutment assemblies underwent a three-week incubation period at 37°C in sterile tubes containing 5 mL of Staphylococcus aureus broth culture. Through the use of Gram stain and biochemical processes, the resultant colonies were recognized. The mean Log10 colony forming unit (CFU) in group A was 8.4 and in group B was 7.2. The variation between both groups was found to be considerable (P < 0.05). Microgap was more in group B compared to group A. Bacteria may infiltrate the small area between the implant and the abutment. Compared to dental implant fixtures with a tri-lobe internal connection, there was a noticeably higher Log10 CFU in dental implant fixtures with an internal hexagonal geometry.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have