Abstract
Arterial compliance and vessel wall dynamics are significant in vascular diagnosis. We present the design of arterial compliance probes for measurement of local pulse wave velocity (PWV). Two designs of compliance probe are discussed, viz (a) a magnetic plethysmograph (MPG) based probe, and (b) a photoplethysmograph (PPG) based probe. The ability of the local PWV probes to consistently capture carotid blood pulse waves is verified by in-vivo trials on few volunteers. The probes could reliably perform repeatable measurements of local PWV from carotid artery along small artery sections less than 20 mm. Further, correlation between the measured values of local PWV using probes and various measures of blood pressure (BP) was also investigated. The study indicates that such arterial compliance probes have strong potential in cuff less BP monitoring.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.