Abstract

A simple near-infrared (NIR) spectrometer with a wavelength range of 898-2130 nm has recently been applied to diagnose Heliotron J plasmas. It adopts a symmetrical crossed Czerny-Turner mount equipped with a thermoelectrically cooled 512 channel InGaAs linear sensor. Reciprocal linear dispersion was deduced to 96.37 nm/mm at the center of the detector. External filters can be inserted into the path of the collection optics to reject second-order spectra, as needed. Absolute intensity calibration was performed together with a visible spectrometer using a tungsten halogen lamp, and the effect of the transmittance fringe in the visible region of the applied long-pass filter on the NIR calibration was investigated. The intended application of the NIR spectrometer includes extending the wavelength region of a spectral monitor to less contaminated regions for Heliotron J plasma studies. In preliminary measurements, we observed the Paschen series for the hydrogen pellet injection plasma and two atomic helium lines, i.e., 2S-2P singlet and triplet lines, in helium gas puffing experiments. A continuum spectrum in this regime that is attributable to black-body radiation from hot spots on the plasma-facing components was identified. In addition, this may also be used to monitor background radiation in the YAG-Thomson scattering signals near 1064 nm.