Abstract
PET detectors with depth-of-interaction (DOI) capability have been studied to improve imaging resolution widely over the world. Since discrete DOI and continuous DOI detection with dual-ended readout technology have their respective limitations, we in this work focus on the continuous DOI detection with single-ended readout using a monolithic LSO scintillator and a multi-pixel photodetector. Based on a non-linear least squares data fitting method and Geant4 simulation, we studied the relationship between the spatial resolution of gamma positioning and the pixel number of photodetector. The results show that for a pixel number larger than 6x6, the positioning spatial resolution does not become significantly better when increasing the pixel number moreover. Another aspect studied is the effect of crystal thickness on the spatial resolution. Increasing the thickness of crystal leads to higher detection efficiency but lower spatial resolution.