Abstract
Energy selecting electron microscopes of the Castaing-Henry prism-mirror-prism design suffer from a loss of image and energy resolution with increasing field of view. These effects can be qualitatively understood by examining the focusing properties of the prism shown in Fig. 1. A cone of electrons emerges from the entrance lens crossover A and impinges on the planar face of the prism. The task of the prism is to focus these electrons to a point B at a focal distance f2 from the side of the prism. Electrons traveling in the plane of the diagram (i.e., the symmetry plane of the prism) are focused toward point B due to the different path lengths of different electron trajectories in the triangularly shaped magnetic field. This is referred to as horizontal focusing; the better this focusing effect the better the energy resolution of the spectrometer. Electrons in a plane perpendicular to the diagram and containing the central ray of the incident cone are focused toward B by the curved fringe field of the prism.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings, annual meeting, Electron Microscopy Society of America
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.