Abstract

Measurements of electrons striking the sample in the Environmental Scanning Electron Microscope (ESEM) are needed to begin to understand the effect of the presence of the gas on analytical measurements. Accurate beam current is important to x-ray microanalysis and it is typically measured with a faraday cup. A faraday cup (Figure 1) was constructed from a carbon block embedded in non-conductive epoxy with a 45 micrometer bore platinum aperture over the hole. Currents were measured with an electrometer and recorded as instrument parameters were varied.Instrument parameters investigated included working distance, chamber pressure, condenser percentage, and accelerating voltage. The conditions studied were low vacuum with gaseous secondary electron detector (GSED) voltage on; low vacuum with GSED voltage off; and high vacuum (GSED off). The base conditions were 30 kV, 667 Pa (5 Torr) water vapor, 100,000x magnification with the beam centered inside aperture, GSED voltage at 370 VDC, condenser at 50%, and working distance at 19.5 mm. All modifications of instrument parameters were made from these conditions.