This paper describes an adjustable loading device that can substitute for the tissue losses of various sized patients in a whole-body MR imager. It resembles a lowpass birdcage resonator but with resistors replacing the capacitors on the cylindrical surface. Power dissipated is a monotonic function of the total surface conductance of the loader. The loader can be used in conjunction with a low loss water filled phantom to measure signal-to-noise ratios (SNR). The impact on measured SNR due to screening by the loader and changing the phantom size or composition are also briefly discussed.