Abstract

Radon is one of the most natural forms of radiation for human exposure. However, high-accuracy measurement of natural radon in water samples is very challenging due to the background correction, data acquisition, and sampling time. Liquid scintillation counter (LSC) and semiconductor alpha spectroscopy detectors are the most commonly used methods of determining radon concentration in water. The present study utilizes both methods to estimate radon in groundwater collected from various locations in the northeast region of Saudi Arabia. The estimated radon concentrations using Hidex 300SL are compared with a Durridge RAD7 detector to evaluate each apparatus’s abilities, advantages, and disadvantages. Both methods show radon concentrations between 0.1 and 3.20 Bq/L with an average of 0.96 Bq/L, with a standard deviation of 0.82 Bq/L. The estimated values are found to be in the safe limit recommended by the USEPA and EAEC and are far below the safe level recommended by UNSCEAR and the WHO. Comparing the estimated radon concentration using the two methods shows that although the two devices have many advantages and disadvantages based on the two different techniques, the experimental results are almost the same with experimental error.