Abstract

Between 1993 and 1997, 21 consecutive revision total hip arthroplasties were performed in 20 patients using hydroxyapatite (HA) granules supported by a Kerboull-type reinforcement acetabular device. Acetabular bone loss according to the American Academy of Orthopaedic Surgeons (AAOS) system was type II for 5 hips and type III for 16 hips. Autografts were used to reconstruct the major segmental defects in 6 hips. The mean follow-up period of the series was 5 years and 4 months. No migration of the socket was seen. Slight inclination of the acetabular device was noted in 1 hip without functional deficits. The mean preoperative Merle d’Aubigné hip functional score was 9.3 versus 15.4 at the latest follow-up evaluation. Acetabular reconstruction with HA granules and a Kerboull-type acetabular device provided satisfactory 3- to 8-year clinical and radiographic results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.