Abstract

This paper presents a comprehensive methodology for the precise calibration of high-voltage capacitance and dissipation factor (DF) bridges. The technique involves meticulous adjustments using a digital high-precision phase and ratio-measuring system to determine correction factors for DF and capacitance measurements. Unlike other methods that are limited to a narrow range, this approach allows calibration across the entire operational working spectrum of any bridge. While the method has been developed up to 1 kHz, its adaptability for future requirements is facilitated by the developed software. This method has been carried out by calibrating a highly accurate bridge for a capacitor ratio of up to 100, with DFs ranging from −0.01 to 0.01 and currents ranging from 0.1 mA to 1 A. Linear correction factors are established, with their accuracies being rigorously quantified. This methodology achieves expanded uncertainties as low as 3.5 µF/F (3.5 ppm) for capacitance and 2.2 × 10−⁶ for DFs, significantly enhancing measurement reliability across diverse high-voltage capacitor applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.