Abstract

Determination of high sulfur content in sulfide minerals is crucial for various applications, including geochemistry, mining, and environmental monitoring. While inductively coupled plasma optical emission spectrometry (ICP-OES) is the preferred method for sulfur analysis, conventional sample preparation techniques often fall short when dealing with high sulfur concentrations. This study investigated optimized digestion conditions for accurate sulfur determination in sulfide minerals using ICP-OES. We employed anti-corrosion acid digestion bombs, an aqua regia–hydrofluoric acid system, and a digestion temperature of 180 °C, achieving exceptional results. Certified reference material recoveries ranged from 99.9 to 100.5%, with relative standard deviations between 0.42 and 0.84%. These results demonstrate the high accuracy and precision of the method. The optimized digestion protocol presented in this study provides a robust and reliable approach for accurate sulfur analysis in sulfide minerals, particularly for high concentrations.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.