Abstract

There is no enough knowledge about the accuracy of intraocular lens (IOL) power calculation formulas in steep corneas. This study may be the first one that compares the accuracy of the SRK II formula with Holladay1, Hoffer Q and Haigis formulas in steep corneas. We reported a case of a 60-year-old female, with a cataract in the left eye and with a steep cornea. We used the modern formulas; Holladay1, Hoffer Q and Haigis. The result (+7.0D) was unexpected compared to the manifest refraction and to the IOL power calculated in the right eye using the same formulas which was (+17.0D). We implanted (+12.0D) Sensar 1-piece IOL depending on our clinical experience. The post-operative refraction was (+0.00/-1.75axis106). Postoperative, we used the patient data to find the best formula in this case. We found that the SRK II (A118) result was (+11.5D) and thus this formula was the most accurate in this case. Keywords: SRK II; Holladay1; Hoffer Q; Haigis

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.