Abstract

PurposeThis study aims to compare the accuracies of intraocular lens (IOL) power calculation formulas when targeting myopia versus emmetropia. MethodsA total of 450 patients were included, with 225 patients targeting emmetropia and 225 patients aiming for approximately −2.0 diopters of myopia. This retrospective analysis utilized data from a single eye of each patient, with preoperative biometric measurements obtained using the IOL Master 700. The study considered established formulas such as Haigis, Hoffer Q, Holladay 1, Holladay 2, and SRK/T, as well as modern formulas including Barrett Universal II, Cooke K6, EVO 2.0, Hill-RBF, Hoffer QST, Kane, Olsen, and PEARL-DGS. Statistical analyses, including Friedman test and post hoc analysis, were employed to compare the accuracy of each IOL power calculation formula between the two groups. Additionally, a multiple regression analysis was conducted to identify variables influencing the accuracy of intraocular lens power calculation formulas. ResultsIn targeting myopia, all IOL formulas tended to exhibit a greater refractive error compared to when targeting emmetropic eyes. Notably, the Haigis, SRK/T, and Holladay 2 formulas were found to be highly influenced by this trend, while the modern formulas were less affected. ConclusionThe accuracy of IOL power calculation formulas diminishes when targeting myopia in comparison to emmetropia. However, the modern formulas appear less susceptible to this trend. Consequently, when aiming for myopia, the use of the modern formulas is recommended for enhanced accuracy in IOL power calculation.