Abstract

The Streaked Optical Pyrometer (SOP) is a visible diagnostic widely used to study the warm dense matter regime at high energy laser facilities, gas guns, or ion accelerators. It is usually coupled with a Velocity Interferometer System for Any Reflector (VISAR) diagnostic for simultaneous shock wave velocity, reflectivity, and temperature measurements to study the Equation of State (EOS) of materials. While VISAR is a well-mastered technology that provides velocity measurements with low relative uncertainties (close to percent), SOP diagnostics still suffer from high imprecision. In this article, we present a new calibration method in order to obtain absolute temperature measurements with reduced uncertainties. This approach is based on a novel light source: a Ce:YAG luminescent concentrator pumped by LEDs. This device produces enough optical power for calibration at the nanosecond sweep duration of the streak camera. As a demonstration, it has first been installed at the LULI facility and tested on quartz samples shocked at temperatures above 4000 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.