Abstract

A stroboscopic pulse radiolysis system is described which detects the transient absorption generated by a 20 psec pulse from a 20 MeV linear accelerator. The time resolution is obtained by using, as the analyzing light, the short pulse of ?erenkov light generated by the portion of the electron beam which passes through xenon. The remaining portion of the electron beam is delayed by passing around a 270° magnet and the delay of the light may be varied by an optical delay line. Block diagrams for the electronics and the optics are given. The magnitude of sources of errors are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.