Abstract

In this paper we report a systematic recovery study based on reversed phase high performance liquid chromatography (RP-HPLC) separation and mass spectrometric (MS) based fractionation. Factors including a compound's physicochemical properties, column mass loading and presence of impurities were investigated through commercially available compounds. Results suggest that the delay time between MS peak detection and fraction collection, fraction detector's signal-to-noise ratio and compound's base peak width in the chromatogram have the biggest impacts on purification recovery. In an effort to assess sample recovery within our high throughput purification process, re-purification was performed on four compound libraries that were synthesized in-house. Reproducible recoveries (>80%) were achieved in all tests.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.