Abstract

A non-optical, pulsed reset technique for use in charge sensitive preamplifiers with Si(Li) X-ray semiconductor detectors of conventional polarity is described. The phenomenon of impact ionization in the field effect transistor is used for charge restoration. The process is controlled via the detector capacitance. The method needs no extra components in the input circuitry. No significant resolution degradation can be detected up to 600 kcps Mn K intensity if following the 5 μs resetting pulse the signal processor is inhibited for 50 μs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.