Abstract
An HPGe detector equipped with a transistor reset preamplifier and readout with a CAEN DT5781 fast pulse digitizer was employed in the measurement of X-rays from kaonic lead at the DAΦNE e+e− collider at the Laboratori Nazionali di Frascati of INFN. A thin scintillator in front of a lead target was used to select kaons impinging on it and to form the trigger for the HPGe detector. We present the results of the kaonic lead feasibility measurement, where we show that the resolution of the HPGe detector in regular beam conditions remains the same as that without the beam and that a satisfactory background reduction can be achieved. This measurement serves as a test bed for future dedicated kaonic X-rays measurements for the more precise determination of the charged kaon mass.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.