Abstract

Background and Objectives: On the basis of newly synthesized i.e. imidazolidine-2tion (IMT), highly selective poly vinyl chloride (PVC) membrane electrode was utilized as a carrier for iodide selective electrode via the incorporation of membrane ingredients within the graphite electrode surface. Method: The impact of different factors consisting of the pH, membrane arrangement and likely intervening anions were examined on the electrode’s response characteristics. The created sensor displayed Nernstian responses to iodide across a concentration of 1× 10-6 to 0.1 M with slopes of 57.8± 0.7 mV per iodide concentration decade and 7 × 10-7M detection limit, across a pH range of 3.0-10.5. Results: The sensor response time was 6 seconds and may be used for a minimum of 2 months with no significant divergence at potential responses. Conclusion: The electrode was implemented with success to determine iodide within synthetic pharmaceutical samples as well as using as an indicator electrode for precipitation titration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.