Abstract

Extrapolation chamber is the only primary standard dosimeter for beta radiation. With the aim to test new configurations and materials using easily-available and low-cost materials and fulfill the need of a chamber for scientific metrological purposes, in this paper the prototype of an extrapolation chamber has been built and its performance has been investigated in the beta radiation field of 90Sr+90Y. The main differences between the chamber and commercially available chambers are the geometry, constituent material and configuration. The obtained results were compared with those of the calibration certificate of the source and an agreement within 4 % was verified. The depth-dose curve was also obtained and compared with the curve published in ISO 6980, showing a good agreement. Moreover, Monte Carlo simulation was undertaken using MCNP4C code and the relative difference of 0.3 % was observed compared to the experiment. All of the results proved the suitability of the chamber in the beta radiation field of 90Sr+90Y.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.