Abstract
We present a novel cryogenic VUV spectrofluorometer designed to characterize wavelength shifters (WLS) crucial for experiments based on liquid argon (LAr) scintillation light detection. Wavelength shifters like 1,1,4,4-tetraphenyl-1,3-butadiene (TPB) or polyethylene naphthalate (PEN) are used in these experiments to shift the VUV scintillation light to the visible region. Precise knowledge of the optical properties of the WLS at liquid argon's temperature (87 K) and LAr scintillation wavelength (128 nm) is necessary to model and understand the detector response. The cryogenic VUV spectrofluorometer was commissioned to measure the emission spectra and relative wavelength shifting efficiency (WLSE) of samples between 300 K to 87 K for VUV (120 nm to 190 nm) and UV (310 nm) excitation. New mitigation techniques for surface effects on cold WLS were established. As part of this work, the TPB-based wavelength shifting reflector (WLSR) featured in the neutrinoless double-beta decay experiment LEGEND-200 was characterized. The WLSE was observed to increase by (54 ± 5) % from room temperature (RT) to 87 K. PEN installed in LEGEND-200 was also characterized, and a first measurement of the relative WLSE and emission spectrum at RT and 87 K is presented. The WLSE of amorphous PEN was found to be enhanced by at least (37 ± 4) % for excitation with 128 nm and by (52 ± 3) % for UV excitation at 87 K compared to RT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.