Abstract

“Coral”-type microstructure carbon films, with and without N 2 incorporation, were grown on sharpened tungsten microprobes by plasma enhanced chemical vapor deposition (PECVD) using H 2/CH 4/N 2 and H 2/CH 4 gas mixtures, respectively. The electrochemical behaviors of the coral-type carbon coated tungsten microprobe, characterized by various concentrations of ferrocyanide in a background of 0.1 M KCl, show excellent structural stability with similar microstructure before and after prolonged analysis without the need of surface pretreatment. The microprobes exhibit quasi-reversible kinetics with high signal-to-noise S/ B ratio. The N 2 incorporated microprobe shows a slightly wider potential window, no surface adsorption of the analyte and higher sensitivity as compared to the sample without nitrogen incorporation. Furthermore, the wide potential window of ∼ 3 V is very good as compared to boron-doped diamond electrodes which are ∼ 3.5 V. This well behaved; broad electrochemical behavior and the simple fabrication method make the “coral” carbon film microprobe an excellent candidate for electrochemical sensing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.