Abstract

The linear CCD camera which applies to micro-image measuring system must have been calibrated before working. The separate calibration method of linear CCD camera’s internal and external orientation parameters is proposed in the paper. The imaging model of the linear camera has been studied, and the equal parallel lines are used to calibrate internal parameters of camera in the experiment where the industrial environment is simulated. In addition, parallel lines and tri-forked stencil are used to calibrate external orientation parameters in the industrial field, such as the attitude angel. The separate calibration method has reduced the calibration task in the industry, and the calibration template is easy to make. The results show that the testing accuracy improves greatly by using the calibrated camera for measuring the gauge, and the calibrated camera can be easily applied to other measurement environment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.