Abstract
A cryogenic micro-calorimeter detector using a small bulk iridium single crystal as both absorber and as transition edge thermometer has been prepared. The detector is voltage biased, and the current through it is measured with a DC-SQUID. In this mode of operation an improvement of the detector performance by the electrothermal feedback is expected. In fact, we observed a stabilization of the working temperature due to this effect. By irradiating the micro-calorimeter with an 241Am X-ray and γ-ray source an RMS energy resolution of σ=150 eV for 14 keV X-rays and σ=250 eV for 60 keV γ-rays was observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.