Abstract
A compact and nearly maintenance-free timing calibration system using a dye laser has been constructed for the KEK TRISTAN TOPAZ barrel TOF counter system. Its timing resolution has been measured to be about 50 ps. The characteristics of its major components have been obtained so as to thoroughly understand their functions and performances.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.