Abstract

An automated ac-dc difference calibration system is described. The system incorporates a new electronic comparator which determines ac-dc differences of thermal voltage converters (TVC's) by simultaneously measuring the difference between nearly equal ac and dc voltages with both the test and standard instruments. The comparator consists essentially of two practically identical channels each containing a digital-to-analog converter balancing circuit, an operational amplifier to amplify voltage imbalance, and an integrator circuit in which a capacitor is charged during an accurately controlled 10-s period. The difference between the ac and dc voltages applied to the test and standard TVC's is computed from differences in capacitor voltages, and the ac-dc difference is derived from the variation between the test and standard channel indications of this voltage difference. Measurements are made in about half the time required for our manual procedures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.