Abstract
Abstract We have developed a new double-pass optical beam deflection instrument for the measurement of diffusion, thermodiffusion and Soret coefficients in liquid mixtures. The increased sensitivity of the instrument results from a second passage of the readout laser beam through the Soret cell containing the sample. An elegant description of the total beam deflection is achieved by means of a transfer matrix formalism. The higher sensitivity allows for a reduction of the length of the detection arm and a compact and stiff design of the instrument. The performance of the new apparatus is demonstrated by its application to polymer analysis for the determination of the molar mass distribution of the polymer from the distribution of diffusion rates by means of the CONTIN algorithm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.