Abstract

A new heating method is proposed to increase the cell temperature of the 6–8 type multi-anvil apparatus without reducing the volume of the sample chamber. The double-layer heater assembly (DHA) has two layers of heaters connected in parallel. The temperature of the cell was able to reach 2500 °C by using 0.025 mm rhenium foils, and the temperature limit was increased by 25% compared with that of the traditional single-layer assembly. The power–temperature relationships for these two assemblies with different sizes were calibrated by using W/Re thermocouple at 20 GPa. When the volume of samples was the same, the DHA not only attained higher temperature, but also kept the holding time longer, compared to the traditional assembly. The results of more than ten experiments showed that the new 10/4 DHA with a relatively large sample size (2 mm in diameter and 4 mm in height) can work stably with the center temperature of the sample cavity exceeding 2300 °C under the pressure of 20 GPa.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.