Abstract

A capillary electrophoresis microchip is developed for the rapid determination of 4-amino-3-methyl- N-ethyl-N-(β-methane sulfonamidoethyl)aniline (CD-3) in commercial colour photographic processing solutions and the applicability of the method is examined. The use of indirect fluorescence as an on-chip detection method is also demonstrated. Using a running buffer at pH 11.9 prepared from disodium hydrogenphosphate and fluorescein the quantitative determination of CD-3 is achieved, resulting in an analysis time of approximately 7 s. Under these conditions, a detection limit of about 5 mg L−1 is obtained, with good linearity between signal and concentration over a range of 5–20 mg L−1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.